McPherson logo
Model 275D Spectrometer

Model 275D Spectrometer Photo - 34kb

The Model 275D is a 0.2 m + 0.2 m focal length double spectrometer system combined in either an additive or subtractive mode. With its two holographic gratings and only two reflective surfaces, it has minimal light loss, maximum stray light rejection and an impressive f / 4.2.

It is an ideal instrument for many spectroanalytical purposes. It features bilaterally adjustable slits, kinematically mounted easy to interchange gratings and the means to attach stepper drives for wavelength scanning.


Specifications for Subtractive Mode Version
Focal Length                                     

0.2 m

Angle between entrance and exit beam    

180 degrees

 f / Number

4.2

Wavelength Range 185 nm to 4000 nm
Grating (G/mm)

1200

600

300

Wavelength Range

-

-

-

From

185 nm

185 nm

185 nm

To

  1000 nm  

  2000 nm  

  4000 nm  

Holographic Range of

Optimal Efficiency (nm)

200 - 800

   400 - 1600     800 - 3200
Optimal Resolution (nm)

0.15

0.30

0.60

Bandwidth (nm)

(250/500/1000/2000 um slits)

(Divide by 2 for Additive Version)

1/2/4/8

2/4/8/16

4/8/16/32

Dispersion (nm/mm)

(Multiply by 2 for Additive Version)

4

8

16

Grating Size

40 mm x 45 mm

Wavelength Accuracy with 1200 G/mm

+/-1 nm

Wavelength Reproducibility 1200 G/mm

+/- 0.1 nm


McPherson, Inc.

7A Stuart Road, Chelmsford, MA 01824

Telephone (978) 256-4512       Fax (978) 250-8625 E-Mail mcp@mcphersoninc.com

Back to the top of the page. Return to the McPherson Homepage.